yy language
    Pulse Vacuum Autoclave
    +
    • Pulse Vacuum Autoclave

    Mechanical-door

    WG series pulsating vacuum sterilizers use saturated steam as the sterilizing medium, by relying on the physical property that the saturated steam can release a large amount of latent heat and moisture during condensation, it renders items to be sterilized in a highly heated and humid state. After a period of heat preservation, the purpose of sterilization is achieved. The effect of cold air on temperature is eliminated by means of pulsating vacuum exhaust, finally the items drying is realized by the vacuum dehumidification combined with jacket.
    WG series pulsating vacuum sterilizers are mainly used in central sterile supply department, third-party disinfecting supply centers, and operating rooms, where medical devices, dressings, rubber, lumen devices, implantable devices, and liquids are sterilized.



    Structure and Characteristic of the Equipment

    Body of the sterilizer: 
    • Designing, manufacturing and inspection are performed according to GB/T 150 "Pressure Vessel", GB 8599 "Technical Requirements for Large Steam Sterilizer Automatic Control Type" and TSG 21-2016 "Stationary Pressure Vessel Safety Technical Supervision Regulations".
    • The main structure is of a rectangular cross section with European ring-type jacket with reinforcing ribs.The doors are pneumatically sealed,and the two doors are interlocked,which fully meets the specific requirements.
    • The materials that come into contact with steam such as the interior liner and door panels are all S30408 stainless steel,with a designed service life up to 10 years and 20,000 sterilization cycles (316L stainless steel is the optional material with designed service life of 15 years and 30,000 sterilization cycles.The gear rack and the teech are made of S30408 stainless steel,and the jacket is made of high-quality Q245R material specific for pressure vessels, with S30408 optional.
    • With standard GMP authentication interface.
    Piping System:
    • 304 stainless steel sanitary pipes, with clamp joints. Adopting GEMU angle seat pneumatic valves imported from Germany, solenoid valves supplied by Taiwan Air TAC and directly connected water ring vacuum pump imported from German SPECK.
    • (Optional) Smart automatic drainage system, water-conservative noise reduction system and pipeline insulation treatment.
    Controlling System:
    • Adopting imported Siemens PLC and colorful touchscreens on the operator's side;the printer prints the process data of each stage in real time manner;with a variety of safety and alarming systems.
    • (Optional) Equipped with a communication interface connected with the quality tracing system of the disinfecting supply center. Intelligent maintenance system,with remote monitoring and maintenance modules.
    • Double pulsation processes more thoroughly remove the nterior air,improving fatigue life of equipment,with excellent drying performance.

     

    Main Parameters and Utilities

    Designed Pressure of Inner Chamber

    -0.1~0.3MPa

    Opening Pressure for Inne Chamber's Safety Valve

    0.28MPa

    Designed Pressure of Jacket

    0.3MPa

    Opening Pressure for Jacket's Safety Valve

    0.28MPa

    Designed Temperature

    150℃

    Precision of Sterilization Temperature Control

    0~2℃

    Maximum Working Temperature

    138℃

    Temperature Uniformity

    ±1℃

    Maximum Working Pressure

    0.25MPa

    Negative Pressure Pulsation Vacuum Amplitude and Times

    Amplitude-80~0KPa, Times 1~99

    Vacuum Leakage Rate

    ≤0.13KPa/min

    Times of Pulses Across Pressure

    Amplitude-80~80KPa, Times 1

    Ultimate Vacuum

    -96KPa

    Times of Positive Pressure Pulses

    1~3


    Size & Utilities 

    Model

    Size of Liner

    Volume (L)

    Overall Size L×W×H(mm)

    Weight (Kg)

    Steam Consumption (Kg/C)

    Tap Water Consumption (Kg/C)

    Power Supply

    Power

    W×H(mm)

    Depth(mm)

    WG-0.25JS(F)

    550×550

    845

    250

    1150×1310×1880

    750

    18

    120

    380V,50HZ

    2KW+(24)KW

    WG-0.36JS(F)

    610×610

    1000

    360

    1320×1350×1930

    850

    22

    180

    380V,50HZ

    2KW+(24)KW

    WG-0.6JS(F)

    610×910

    1150

    600

    1470×1410×2100

    1250

    35

    320

    380V,50HZ

    3KW+(36)KW

    WG-0.8JS(F)

    610×910

    1490

    800

    1820×1410×2100

    1350

    47

    400

    380V,50HZ

    3.5KW+(48)KW

    WG-1.0JS(F)

    610×910

    1815

    1000

    2150×1410×2100

    1550

    55

    500

    380V,50HZ

    3.5KW+(48)KW

    WG-1.2JS(F)

    680×1180

    1500

    1200

    1820×1490×2050

    1650

    65

    600

    380V,50HZ

    3.5KW+(72)KW

    WG-1.5JS

    680×1180

    1900

    1500

    2220×1490×2050

    1850

    80

    750

    380V,50HZ

    4KW

    WG-2.0JS

    900×1380

    1650

    2000

    1980×1750×2160

    2300

    106

    1000

    380V,50HZ

    4.5KW

    WG-2.5JS

    900×1392

    2000

    2500

    2350×1750×2160

    2700

    132

    1250

    380V,50HZ

    6KW

    WG-3.0JS

    900×1380

    2494

    3000

    2830×1860×2160

    3500

    160

    1500

    380V,50HZ

    8KW

    WG-4.0JS

    900×1380

    3250

    4000

    3580×1860×2160

    4500

    215

    2000

    380V,50HZ

    8KW

    Note: (24)(36)(48)(72)represent the power of the built-in steam generator(optional) 

    Mechanical-door

    Pulse Vacuum Autoclave

    WG series pulsating vacuum sterilizers use saturated steam as the sterilizing medium, by relying on the physical property that the saturated steam can release a large amount of latent heat and moisture during condensation, it renders items to be sterilized in a highly heated and humid state. After a period of heat preservation, the purpose of sterilization is achieved. The effect of cold air on temperature is eliminated by means of pulsating vacuum exhaust, finally the items drying is realized by the vacuum dehumidification combined with jacket.

    Pulse Vacuum Autoclave
    +
    • Pulse Vacuum Autoclave

    Send us an email

    sales@jibimed.com

    postmaster@chengyi-cn.com



    Keywords : Pulse Vacuum Autoclave

    Production Control

    With a 30,000 square meter cleanroom, we conduct weekly testing of air settle plate counts, surface microbial colonies, and disinfection of personnel's hands to maintain a clean production environment. Through these measures, we ensure safety and high quality standards.

    Production Control
    Production Control
    Production Control
    Production Control

    Manufacturing Advantages

     With a strong focus on research and development

    Maintains stringent quality control processes throughout the production cycle

    Efficient Management and Excellent Service

    We Are Always Happy To Help

    We are dedicated to providing superior products and services to all our valued customers, in gratitude for their trust and support.

    Submit
    %{tishi_zhanwei}%
    图片名称

    Empowering Healthcare Providers Worldwide


    Tel:+86-510-86270699       +86-510-86296768


    Add:No 38, Yungu Road, Changshou Community, Zhouzhuang Town, Jiangyin City, Jiangsu Province


    Copyright Jiangyin Binjiang Medical Equipment Co., Ltd. All Rights Reserved.