Mobile
Copyright:Jiangyin Binjiang Medical Equipment Co., Ltd. 苏ICP备05069341号-1 Powered By www.300.cn
Page view:
产品名称
WG Series Pulse Vacuum Autoclave
副标题
Fully Automatic with hinge door
WG series pulsating vacuum sterilizers are mainly used in disinfecting supply rooms of hospitals, third-party disinfecting supply centers, and operating rooms, where medical devices, dressings, rubber, lumen devices, implantable devices, and liquids are sterilized.
Product description
规格参数
This scries of pulsation vacuum sterilizer utilizes saturated steam as its sterilization medium which releases a large amount of heat in the stage of condensation, and ensure the sterilizing environment in high temperature and at a certain humidity.
After a period of insulation in such status, sterilization purpose is achieved. The application of pulsating vacuum exhausting has successfully eliminated the temperature influence from cold air, and helping the sterilization items to be dried by means of vacuum and dehumidification between the jacketed layers.
Structure and Characteristic of the Equipment
Body of the strilizer.
It is designcd and manufactured in accordance with GB150 *the Pressure Vessel. GB8599 The Technieal Requirements of Large Steam Serilier-Automatic Control Protocol". and TSG Rt004 the -Safety and Technology Specifieation for Fixcd Monitor VesseNi and related insetion procedures.
The chamber in horizontat rectangular shape with doublelayer structure is made from S3008
Doot sealing: Pneumatic scaling: both doors are interlocked and completely conform to the requirementsof the
specification
With standard GMP authentication interface.
Strlizer piping sysem: the pipeline is designed with the best configuration. All the spares are from qualified brands at homc and abroad.
Sterilizer control system: SIEMENS PLC and SIEMENS touch screen, this system is easy to implement thc program selection, parameter seting equipment opeation, report processing and other functions. It has prfect safety measures The FO valuc and thc temperature time gives doublc guarantee for selizaion A single parameter control is also available. It hasa prfect sterilization records.
Main Parameters and Utilities
Design Pressure |
0.245Mpa |
Working Pressure |
0.225Mpa |
Design Temperature |
139℃ |
Working Temperature |
105-134℃ |
Vacuum |
-0.09Mpa |
Temperature Equilibrium |
≤±1℃ |
Water Source Pressure |
0.15-0.3Mpa |
Compressed Air Pressure |
0.3-0.7Mpa |
Steam Pressure |
0.3-0.7Mpa |
Power |
AC220V or 380V, 50HZ |
Size & Utilities
Model |
Overall Size L×W×H(mm) |
Chamber Size L×W×H(mm) |
Steam Consumption(KG) |
Tap Water Consumption(KG) |
Power(KW) |
Net Weight(KG) |
WG-0.1 | 950×1200×1750 | 560×400×450 | 14KG | 25KG | 2+12KW | 700KG |
WG-0.15 | 1050×1200×1750 | 750×400×450 | 14KG | 25KG | 2+12KW | 800KG |
WG-0.2 | 1120×1305×1840 | 1000×450×450 | 16KG | 30KG | 2+24KW | 850KG |
WG-0.25 | 1100×1350×1900 | 800×600×600 | 18KG | 35KG | 2+24KW | 900KG |
WG-0.36 | 1300×1350×1900 | 1000×600×600 | 25KG | 40KG | 2+24KW | 1000KG |
WG-0.6 | 1500×1360×1950 | 1200×610×910 | 30KG | 45KG | 3+36KW | 1400KG |
WG-0.8 | 1800×1360×1950 | 1500×610×910 | 40KG | 50KG | 3KW | 1600KG |
WG-1.2 | 1750×1440×1950 | 1450×680×1180 | 48KG | 55KG | 4KW | 1800KG |
WG-1.5 | 2150×1440×1950 | 1850×680×1180 | 55KG | 60KG | 4KW | 2000KG |
WG-2.0 | 1950×1750×2200 | 1600×900×1400 | 65KG | 65KG | 4.5KW | 2500KG |
WG-2.5 | 2300×1750×2200 | 1950×900×1400 | 75KG | 75KG | 6KW | 3000KG |
WG-3.0 | 2700×1750×2200 | 2400×900×1400 | 90KG | 90KG | 8KW | 3500KG |
Note: 2436 represent the power of the built-in steam generator (optional)
Previous article
无
WG Series Pulse Vacuum Autoclave
Next article
More Products +
Vertical Pressure Steam Sterilizer
副标题
ELECTRIC-HEATED VERTICAL STEAM STERILIZE (hand round automatic)
副标题
WG Series Pulse Vacuum Autoclave
副标题
Fully Automatic with hinge door
WG Series Pulse Vacuum Autoclave
副标题
Fully Automatic with sliding door
Page up
1
2
...
7