    Use of the Pressure Cooker

    Release time:2015-09-23

    High temperature pot How to use
    1. First, remove the inner bucket, then add an appropriate amount of water to the outer pot, ensuring that the water level is flush with the triangular rack.
     
    2. Place the items back into the sterilization bucket and load them carefully, ensuring they are not packed too tightly to allow free steam circulation and maintain effective sterilization. Neither the neck of the Erlenmeyer flask nor the opening of the test tube should come into contact with the bucket walls, as condensation could drip onto the wrapping paper at the bag opening and penetrate the cotton plug.
     
    3. Install the lid and insert the vent hose on the lid into the vent groove of the inner bucket. Then, tighten the two opposing bolts simultaneously in a symmetrical, pair-by-pair manner to ensure even tension and prevent air leakage.
     
    4. Heat the autoclave using an electric heater or a gas burner, and simultaneously open the exhaust valve to bring the water to a boil, thereby expelling the cold air from the chamber. Once all the cold air has been completely vented, close the exhaust valve and allow the temperature inside the chamber to rise gradually as the steam pressure increases. When the chamber pressure reaches the desired level, reduce the heat source and maintain that pressure for the required duration. In this experiment, the settings are 1.05 kg/cm², 121.3°C, and 20 minutes. 5. After the prescribed time has elapsed, disconnect the power supply or turn off the gas, and allow the chamber temperature to drop naturally. When the pressure indicated by the gauge falls to zero, open the exhaust valve, loosen the bolts, lift the lid, and remove the items. If the pressure has not yet dropped to zero when the exhaust valve is opened, the sudden release of pressure inside the chamber will cause the culture medium to be forced out of the flask or test tube due to the resulting pressure imbalance between the inside and outside of the container, leading to contamination of the cotton plugs with the medium.
     
    6. Place the retrieved culture medium in a 37°C incubator for 24 hours; if no contaminating microorganisms are observed upon inspection, it is ready for use.
     
    However, there are a few points to keep in mind: 1. The pot must contain sufficient water. 2. The pot must be completely airtight.
     
    3. After use, be sure to allow it to reach the appropriate temperature before opening.

    + More Products

    2025-05-19

    The 2025 Shanghai Pujiang International Intellectual Property Forum Concludes Successfully, with Jiangyin Binjiang Medical Equipment Co., Ltd.

    The 2025 Shanghai Pujiang International Intellectual Property Forum Concludes Successfully, with Jiangyin Binjiang Medical Equipment Co., Ltd.

    Learn More →

    2021-06-09

    Solemn Statement

    Recently, Jiangyin Binjiang Medical Equipment Co., Ltd. (hereinafter referred to as “our company”) discovered that information about our products and promotional materials are being posted on Taobao and various other online platforms. We hereby solemnly declare that, apart from our official website, our company has never registered any merchant accounts on any other online platforms, nor has it ever published product information or promotional materials on such platforms.

    Learn More →

    2020-11-25

    How to Properly Use Electric Heating Equipment Such as Oven Dryers

    The oven must be placed indoors in a dry, level location to prevent vibration and corrosion. Always ensure that the power supply is properly monitored: select a power switch and wiring that match the oven’s power consumption and capacity, and provide a reliable grounding connection.

    Learn More →

    2020-11-16

    Precautions for Steam Sterilizers

    Learn More →

    图片名称

    Empowering Healthcare Providers Worldwide


    Tel:+86-510-86270699       +86-510-86296768


    Add:No 38, Yungu Road, Changshou Community, Zhouzhuang Town, Jiangyin City, Jiangsu Province


    Copyright Jiangyin Binjiang Medical Equipment Co., Ltd. All Rights Reserved.